Art.Nr.: GMH3431Produktbeschreibung
conductivity measuring device incl. 2-pole measuring cell
- display of resistivity, salinity or TDS (total dissolved solids)
- in accordance to the regulations of the drinking water ordinance (TrinkwV 2001) and DIN EN 27888
measuring ranges: | conductivity: 0.0 μS/cm to 200.0 μS/cm 0 μS/cm to 2,000 μS/cm 0.00 μS/cm to 20.00 mS/cm 0.0 μS/cm to 200.0 mS/cm manually selectable or AutoRange temperature: -5.0°C to +100.0°C resistivity: 0.005 kOhm * cm to 100.0 kOhm * cm salinity: 0.0 g/kg to 70.0 g/kg water TDS: 0 mg/l to 1,999 mg/l |
accuracy (±1 digit) (at nominal temperature = +25°C): | conductivity: ±0.5 % of m.v. ±0.3 % FS or ±2 μS/cm temperature: ±0.2 % of m.v. ±0.3 K |
cell correction: | adjustable 0.800 cm-1 to 1.200 cm-1 manually or automatically with selectable reference solution |
temperature compensation: | automatical or disengageable, by temperature sensor integrated in the electrode |
type of compensation: | nLF: non-linear function of natural waters acc. to DIN EN27888 (ISO 7888) (reference temperature selectable: +20°C or +25°C) Lin: linear compensation from 0.3 %/K to 3.0 %/K (reference temperature selectable: +20°C or +25°C) off: no compensation |
display: | two 4 digit LCD displays (12.4 mm or 7 mm high) for conductivity (resistivity, salinity, TDS) and temperature, or for min, max value, hold function, etc. and additional indicator arrows |
measuring cell: | conductivity measuing cell with integrated temperature sensor in shaft, electrode material: graphite. shaft material: PPE, PS (GMH 3431), the graphite electrodes are an excellent solution for sewage and can be cleaned easily, GMH 3431: 2-pole; GMH 3451: 4-pole |
warranty for sensor element: | 12 months |
working conditions: |
device: -25°C to +50°C, 0 % to 95 % r. h. measuring cell: -5°C to +80°C (permanent) to +100°C (short-term) |
relative humidity: | 0 % to +95 % r.h. (non-condensing) |
interface: | serial interface, connectable to RS232 or USB interface of a PC via electrically isolated interface converter GRS 3100, GRS 3105 or USB 3100 N (accessories) |
control elements: | 6 membrane keys for ON/OFF-switch, selection of meas. range, min and max value memory, hold-function, etc. |
power supply: | 9 V battery as well as additional power supply unit (1.9 mm internal pin Ø) for external 10.5 V to 12 V DC supply (suitable power supply: GNG10/3000) |
battery life: | approx. 150 h |
housing: | impact-resistant ABS plastic housing, membrane keyboard, transparent panel, integrated pop-up clip |
dimensions: | device: 142 x 71 x 26 mm (H x W x D) electrode shaft: approx. 120 mm long, Ø approx. 12 mm, 1 m permanently fixed connection cable between electrode and device |
weight: | approx. 230 g (incl. battery and measuring cell) |
catalogue page as .pdf:
instruction manual as .pdf:
Empfohlenes Zubehör | Weitere Artikel
Geräte-Schutztasche | ST-R152,30 EUR |
|
Koffer | GKK 300028,30 EUR |
|
Steckernetzgerät | GNG 10/300071,60 EUR |
|
Schnittstellen-Konverter | GRS 310093,10 EUR |
|
USB Schnittstellen-Konverter | USB 3100N93,10 EUR |
|
Software für Messdatenerfassung | EBS 20M / 60M121,00 EUR |
|
ISO-Kalibrierschein Leitfähigkeit | WPL 10243,00 EUR |
|
ISO-Kalibrierschein Leitfähigkeit | WPL 3155,00 EUR |
|
ISO-Kalibrierschein Leitfähigkeit, für Reinstwasser | WPL 3-RW207,00 EUR |
|
Leitfähigkeitskontrolllösung | GKL 1009,15 EUR |
|
Geräte-Schutztasche | ST-R1-US55,30 EUR |
|
Koffer | GKK 300155,30 EUR |